1.1This procedure covers the determination of selected polyfluorinated compounds (PFCs) in a soil matrix using solvent extraction, filtration, followed by liquid chromatography (LC) and detection with tandem mass spectrometry (MS/MS). These analytes are qualitatively and quantitatively determined by this method. This method adheres to multiple reaction monitoring (MRM) mass spectrometry. This procedure utilizes a quick extraction and is not intended to generate an exhaustive accounting of the content of PFCs in difficult soil matrices. An exhaustive extraction procedure for polyfluoralkyl substances, such as published by Washington et al.,2 for difficult matrices should be considered when analyzing PFCs.
1.2Units-The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.
1.3The Method of Detection Limit3 and Reporting Range4 for the target analytes are listed in Table 1.
Analyte
MDL
(ng/kg)
Reporting Limit
(ng/kg)
PFTreA
6.76
25-1000
PFTriA
5.26
25-1000
PFDoA
3.56
25-1000
PFUnA
2.45
25-1000
PFDA
5.54
25-1000
PFOS
18.83
50-1000
PFNA
2.82
25-1000
PFecHS
2.41
25-1000
PFOA
6.24
25-1000
PFHxS
7.75
25-1000
PFHpA
5.80
25-1000
PFHxA
15.44
50-1000
PFBS
6.49
25-1000
PFPeA
20.93
125-5000
PFBA
22.01
125-5000
FHEA
199.04
600-20 000
FOEA
258.37
750-20 000
FDEA
137.46
500-20 000
FOUEA
4.85
25-1000
FhpPa
5.09
25-1000
FHUEA
3.50
25-1000
(A)Abbreviations are defined in 3.2.Analyte/Surrogate
LV1
LV2
LV3
LV4
LV5
LV6
LV7
LV8
LV9
PFPeA, PFBA
25
50
100
200
300
400
500
750
1000
PFTreA, PFTriA, PFDoA, PFUnA, PFDA, PFOS, PFNA, PFHxA, PFHpA, PFBS, PFechS, PFOA, PFHxS, FOUEA, FHUEA, FHpPA, MPFBS, MPFHxA, MPFUnA, MPFOA, MPFDA, MPFOS, MPFNA, MPFHxS, MPFBA
5
10
20
40
60
80
100
150
200
FHEA, FOEA, FDEA
100
200
400
800
1200
1600
2000
3000
4000
1.3.1The reporting limit in this test method is the minimum value below which data are documented as non-detects. Analyte detections between the method detection limit and the reporting limit are estimated concentrations and are not reported following this test method. In most cases, the reporting limit is calculated from the concentration of the Level 1 calibration standard as shown in Table 2 for the polyfluorinated compounds after taking into account a 2-g sample weight and a final extract volume of 10 mL, 50% water/50% MeOH with 0.1% acetic acid. The final extract volume is assumed to be 10mL because 10 mL of 50% water/50% MeOH with 0.1% acetic acid was added to each soil sample and only the liquid layer after extraction is filtered, leaving the solid and any residual solvent behind. It is raised above the Level 1 calibration concentration for PFOS, PFHxA, FHEA, and FOEA; these compounds can be identified at the Level 1 concentration but the standard deviation among replicates at this lower spike level resulted in a higher reporting limit.
1.4This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health and environmental practices and determine the applicability of regulatory limitations prior to use.
1.5This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.